Research Systems

12 21 , 2020

Medicell spore & virus free water system for microbiological research and cellular cultivation

A complete system of Pre-treatment-Distillation-Storage-Distribution-Dispensing where all parts in contact with the distillate are of borosilicate, titanium or PTFE.

The produced distillate is free of bacteria, toxins, virus, spores, DNAs and RNAs.

The system is completely closed from the incoming feed water up to the bottle or container to be filled. This is achieved by a special distribution and dispensing system ensuring absolute cleanness and sterility. The distillate is stored and distributed at ambient temperature.

A special washing machine is also supplied with the equipment to ensure that the bottles or objects to be filled are sufficiently clean before the water is filled.

Several years of testing at an independent laboratory has given excellent research results reducing spontaneous infections and ensuring an even growth rate in the cultivations.

DESIGN FEATURES

* Closed system up to the user points.
* Unique quality assurance system.
* Unique dispensing system.
* Storage & distribution at ambient….temperature.
* Regular sterilisation not required.
* Easy to install.

AVAILABLE SYSTEM CAPACITIES, electrically heated

E-8-C, system of 8 L/h with max storage tank size of 200 L

E-25-C, system of 25 L/h with max storage tank size 600 L

E-50-C,system of 50 L/h with max storage tank size 1200 L.

PLC AUTOMATION

* Fully automatic function.
* Automatic start/stop on signal from…the storage tank.
* Automatic user point regulation.
* Discharge of sub-standard distillate during… start,operation and stop.
* Automatic surveillance & alarms of critical… process values.

FACTORY ACCEPTANCE TESTING

All AQUA-NOVA stills are production FAT tested at our factory before delivery. The tests are documented and the produced distillate quality is verified by an independent laboratory.

The software is developed as per GAMP specifications.

When ordered, full Pre IQ and OQ documentation will be supplied with the machines, including calibration documents.

The standard delivery documentation consists of:

* Authorised drawings with material and weld…certificates.
* Mechanical component documentation and…manuals.
* Electrical and instrument documentation and… manuals.
* Machine installation, operation and service… manual

Other News

12 21 , 2020

Aqua-Nova Services

The Water Still is the heart and the WFI distribution the bloodstream of your sterile, liquid production. When the Still together with the storage vessel and distribution loop is maintained in good health the risk to product contamination or eventual roduction halts will be avoided. There […]

12 21 , 2020

RO-Systems

Purified Water RO Unit, double or single in combination with CEDI The units have a number of design features to achieve an optimised function and a safe water quality exceeding the latest editions of USP and EP for Purified Water (PW) or Highly Purified Water […]

12 21 , 2020

Laboratory Still

Laboratory “Water For Injection” Still and Pure Steam Generators The units have a number of design features to achieve an optimised function and a safe distillate quality exceeding the latest editions of USP and EP for Water For Injection. They characterised by it´s reliability and […]

12 21 , 2020

Pure Steam Generators

3rd Generation Pure Steam Units with Pure Steam of WFI qualityThe units have a number of design features to achieve an optimised function and a safe Pure Steam quality, condensed as per the latest editions of USP and EP for Water For Injection. DESIGN FEATURES […]

12 21 , 2020

Single-effect WFI Stills

The units have a number of design features to achieve an optimised function and a safe distillate quality exceeding the latest editions of USP and EP for Water For Injection. DESIGN FEATURES * Fully automatic function and surveillance. * Unique 4 step separation system. * […]

12 21 , 2020

Combis TWO in ONE

TWO in ONE Pure Steam Production! The increasing demand for flexibility calls for new technical solutions concerning the  possibility of producing Pure Steam and WFI at the same time. All multi-effect stills have a limited possibility of producing Pure Steam from the 1st column. Normally […]